6 Potentiometric measurement of pH
Theory:
see Electrochemical cells and electrodes, p. 6-29
Potentiometric measurement of pH using hydrogen electrode, p. 7-36
Potentiometric measurement of pH using glass electrode, p. 8-41, in
Kopecký F., Kaclík P., Fazekaš T.:
Laboratory manual for physical chemistry,
Department of Physical Chemistry, FaF UK, Bratislava, 1996

Task:
Measure pH of 7 different aqueous solutions using:
A: hydrogen electrode
B: glass electrode.
C: quinhydrone electrode
Equipment and chemicals:
Chemicals: 7 aqueous solutions with different pH, distilled water
Equipment for task A:
The potentiometer for the measurement of the electromotoric force, cell with the hydrogen and calomel electrodes, hydrogen gas in the steel cylinder with valve and manometers, thermometer, beakers, filtering paper

An electrochemical cell (Fig. 7-1, p. 7-38) includes: hydrogen electrode with the platinized platinum (1) placed inside of tube connected with the hydrogen gass valve
(1). Saturated calomel electrode (3) is used as the reference electrode. Electrodes are immersed in the measured solution (2). For storage, hydrogen electrode should be kept in distilled water, and the calomel electrode in the saturated KCl solution. It is necessary to prevent the sensitive parts of electrodes from drying out.
Procedure:
Manipulation with the hydrogen gas valve is allowed to the instructor only, or under his supervision!
1. Take off electrodes from the storage solution (shift them vertically up without leaving the holder). Rinse them with distilled water, and dry gently with filtering paper.
2. Fill the beaker with measured solution.
3. Put the electrodes to the solution in the beaker. At least one half of the platinum plate of the hydrogen electrode must be immersed in the measured solution. 1

4. Connect the electrodes with the potentiometer. Remember, the hydrogen electrode is the negative terminal, and the calomel electrode is the positive terminal. 5. Ask the teacher to check the set up, and to set a flow of the hydrogen. The hydrogen gass now bubbles gently through the measured solution.
6. To saturate the solution with the hydrogen takes about 10-15 minutes.
7. Read the electromotoric force (E in mV) from the display of potentiometer when its value is stable within 0.5 mV.
8. Stop the flow of hydrogen with the valve (1).
9. Write the E value down to the Tab. 1.
10. Use part of the solution for experiment B. Give back the rest of the solution to the bottle with the original solution.
11. Repeat the procedure for each solution.

Data treatment:
Express the pH of the solutions using data from the Tab. 1 according to the relation: pH =

E − Ekal
2.303( RT / F )

(1)

where E is the measured electromotoric force (in V), Ekal is the potential of the calomel electrode at given temperature (see Tab.2). R=8.314 Jkmol-1 is the gas constant, F =96485 Cmol-1 is the Faraday constant, T is the temperature. The values of 2.303(RT/F) as a function of temperature are given in the Tab.2.

Equipment for task B:
The pH meter with the pH probe, beakers

The pH probe consists from:

1. a sensing part of electrode, a bulb made from a specific glass
2. sometimes the electrode contains a small amount of AgCl precipitate inside the glass electrode
3. internal solution, usually 0.1M HCl for pH electrodes
4. internal electrode, usually silver chloride electrode or calomel electrode 5. body of electrode, made from non-conductive glass or plastic.
6. reference electrode, usually the same type as 4
7. junction with studied solution, usually made from ceramics or capillary with asbestos or quartz fiber.

2

A typical modern pH probe is a combination electrode, which combines both the glass and reference electrodes into one body. The bottom of a pH electrode balloons out into a round thin glass bulb. The pH electrode is best thought of as a tube within a tube. The inside most tube (the inner tube) contains an unchanging saturated KCl and a 0.1M HCl solution. Also inside the inner tube is the cathode terminus of the reference probe. The anodic terminus wraps itself around the outside of the inner tube and ends with the same sort of reference probe as was on the inside of the inner tube.
Both the inner tube and the outer tube contain a reference solution but only the outer tube has contact with the solution on the outside of the pH probe by way of a porous plug that serves as a salt bridge.
This device is essentially a galvanic cell. The reference end is essentially the inner tube of the pH meter, which for obvious reasons cannot lose ions to the surrounding environment (as a reference is good only so long as it stays static through the duration of the measurement). The outer tube contains the medium, which is allowed to mix with the outside environment (and as a consequence this tube must be replenished with a solution of KCl due to ion loss and evaporation).
The measuring part of the electrode, the glass bulb on the bottom, is coated both inside and outside with a ~10nm layer of a hydrated gel. These two layers are separated by a layer of dry glass. The silica glass structure is shaped in such a way that it allows Na+ ions some mobility. The metal cations (Na+) in the hydrated gel diffuse out of the glass and into solution while H+ from solution can diffuse into the hydrated gel. It is the hydrated gel, which makes the pH electrode an ion selective electrode. H+ does not cross through the glass membrane of the pH electrode, it is the Na+ which crosses and allows for a change in free energy. When an ion diffuses from a region of activity to another region of activity, there is a free energy change and this is what the pH meter actually measures. The hydrated gel membrane is connected by
Na+ transport and thus the concentration of H+ on the outside of the membrane is
'relayed' to the inside of the membrane by Na+.
Procedure:
The fragile glass electrode must be handled with care, it should be rinsed carefully before and after the use. Between experiments, electrode should be kept in the storage solution. 1.

2.
3.
4.
5.
6.
7.

Take off the electrode from the storage solution (shift it vertically up without leaving the holder). Rinse it with distilled water, and dry gently with filtering paper.
Fill the beaker with measured solution.
Put the electrode back to the beaker. The bulb and also the junction of the reference electrode (7) must be immersed in the solution.
Read the pH from the display of pH meter when the value is stable.
Write the pH value down to the Tab.1
Repeat the procedure for each solution.
Do not forget to write down the laboratory temperature.

3

C: Measurement of pH using quinhydrone eletrode
The quinhydrone electrode is a type of redox electrode which can be used to measure the hydrogen ion concentration (pH) of a solution in a chemical experiment.
It provides an alternative to the commonly used glass electrode in a pH meter.
Redox reaction
It is familiar from introductory chemistry that oxidation is the removal of electrons from a species, a reduction is the addition of electrons to a species, and a redox reaction is a reaction in which there is a transfer of electrons from one species to another. The electron transfer may be accompanied by other events, such as atom or ion transfer, but the net effect is electron transfer and hence a change in oxidation number of an element. The reducting agent (or reductant) is the electron donor, and oxidizing agent (or oxidant) is the electron acceptor. It should be also familiar that any redox reaction may be expressed as the difference of two reduction half-reactions, which are conceptual reactions showing the gain of electrons. The reduced and oxidized species in a half-reaction form a redox couple. In general we write a couple as Ox/Red:
Reduction: Oxz + ne- → Redz-n (oxidation number decreases)
Oxidation: Redz-n → Oxz + ne- (oxidation number increases)
What can be written as:
Ox z + ne − ← Re d z − n
→

(2)

Redox electrodes
A redox electrode consists of an inert material like platinum or gold dipping into a solution containing a chemical species in two different oxidation states. The transfer of electrons between the species takes place through the inert material. The electrode potential is given by

E = E0 +

[a ]
RT
ln ox nF [aRe d ]

(3)

where E0 is the standard potential of electrode, R the gas constant, T (K) the temperature, F the Faraday constant, n is a number of electrons in the electrode reaction, aox and aRed are acivities of the oxidant and reductant, respectively.

Quinhydrone electrode
The quinhydrone electrode consists from a platinum dips into a solution saturated with quinhydrone. Quinhydrone (HQ) is a slightly soluble compound formed by the combination of one mole of quinone (Q) and one mole of hydroquinone (H2Q):

4

The electrode reaction is

C 6 H 4 O2 + 2 H + + 2e − ← C 6 H 4 (OH ) 2
→
Quinone (Q)

(4)

Hydroquinone (H2Q)

Quinone is oxidant, and hydroquinone is reductant in this reaction. This electrode is very easy to prepare and handle. Pure solid quinhydrone is dissolved in the solution to be measured until the solution is saturated and excess is present. A platinum wire is dipped in this solution.
The electrode potential is given by
EQ / H 2 Q = E

0
Q / H 2Q

[ ]
[ ]

aQ ⋅ [a H ]
RT
+ ln 2F a H 2Q

2

(5)

Quinone (Q) and hydroquinone (H2Q) are obtained by dissolving quinhydrone in solution, therefore aH2Q=aQ. Applying the following substitution ln[a H ] = 2.303 ⋅ 2 ⋅ log[a H ] = 2.303 ⋅ (−2). pH
2

(6)

the equation (5) is in form:
0
EQ / H 2Q = EQ / H 2Q − 2.303

RT
⋅ pH
F

(7)

The standard potential of the quinhydrone electrode is given by
0
EQ / H 2Q = −0.6994 + 7.36 ×10 −4 (t − 25)

(8)

where t is the temperature in degree Celsius.
For the potentiometric measurements we combine the quinhydrone electrode with suitable reference electrode to create electrochemical cell. We use a saturated calomel electrode as the reference, and the cell is set up:
Hg Hg 2 Cl 2 KCl (satur.solution ) H 2Q, Q, H + (measur.solution ) Pt
According to Eq.6th.1 (p.6-32), the electromotoric force (E) of the cell is given by

5

E = EQ / H 2Q − E kal

(9)

where EQ/H2Q is the potential of quinhydrone electrode. Finally, from Eq. 7 and 9, we get pH of solution as pH =

0
EQ / H 2Q − E kal − E

(10)

2.303RT / F

The quinhydrone electrode cannot be used in solutions that would react with quinone or hydroquinone. Hydroquinone being a weak acid, the electrode cannot be used above pH 8.5 when the dissociation of hydroquinone becomes appreciable. Another drawback is that quinone is oxidized by air in strongly alkaline medium. In spite of all this, the quinhydrone electrode is frequently used in titration acids with bases since the end point is reached below pH 8 in most cases.

Equipment for task C:
The potentiometer for the measurement of the electromotoric force, cell with the quinhydrone and calomel electrodes, quinhydrone, thermometer, beakers, filtering paper Procedure:
1. Take off electrodes from the storage solution. Rinse them with distilled water, and dry gently with filtering paper.
2. Fill the beaker with measured solution. Approximate values pH of solutions are written on the flask.
3. Take a little bit quinhydrone (black powder) using a wood spatula, and add to the measured solution. Mix the solution carefully using a glass stick to obtain slightly yellow solution saturated with quinhydrone.
4. Put the electrodes to the solution in the beaker.
5. Connect the electrodes with the potentiometer. The quinhydrone electrode is the positive terminal, and the calomel electrode is the negative terminal.
6. Read the electromotoric force (E in mV) from the display of potentiometer when its value is stable within 0.5 mV.
7. Write the E value down to the Tab. 1.
8. Using Eq. (8) calculate the standard potential of quinhydrone electrode at laboratory temperature.
9. Calculate pH of the solution using Eq. (10) and Tab.1
10. Repeat the procedure for each solution where the use of the quinhydrone electrode is possible.

6

Tab. 1. Experimental data
Solution
pH
(approx.value)

Hydrogen electrode

E (mV)

pH

Quinhydrone electrode E (mV)

Glass electrode pH pH

Tab. 2. The potential of the calomel electrode and the value 2.303(RT/F) as a function of temperature. t (oC)
15
16
17
18

Ekal
2.303(RT/F)
t
(V)
(V)
(oC)
0.2473
0.05717
19
0.2468
0.05737
20
0.2462
0.05757
21
0.2456
0.05777
22

Ekal
2.303(RT/F)
t
(V)
(oC)
(V)
0.2450
0.05797
23
0.2443
0.05817
24
0.2436
0.05837
25
0.2430
0.05856
26

Ekal
2.303(RT/F)
(V)
(V)
0.2423
0.05876
0.2417
0.05896
0.2410
0.05916
0.2404
0.05936

The report must include:


Theory (electrode potential, Nernst equation, pH)



Equipment and chemicals



Working procedure and measurements



Tables of results and calculation



In conclusions, compare the pH values obtained by using all methods. The pH

value given on the bottle is only approximate.
Used literature:
Kopecký F., Kaclík P., Fazekaš T.: Laboratory manual for physical chemistry,
Farmaceutical faculty of Comenius University, Bratislava, 1996
J. Oremusová, Manual for laboratory practice in physical chemistry for students of pharmacy, Department of Physical Chemistry, Faculty of Pharmacy, Comenius
University, Bratislava, 2007, in Slovak http://en.wikipedia.org/wiki/PH_glass_electrode Athawale V.D., Mathur P.: Experimental Physical Chemistry, New Age International (P)
Ltd., New Delhi, India, 2001
Manual written by Doc. RNDr. D. Uhríková, CSc.

7…...

Similar Documents

Premium Essay

Ph Measurment

...of paper known as pH paper for very broad measurements of pH. pH paper can be bought for acidic neutral and alkaline compounds. Once dipped into the hair product, the paper changes color depending on the pH of the product. Then the color can be matched to a key (included with the pH paper) to determine the approximate pH. 2 Use a digital pH meter for more precise measurements. Though the specifics of using a pH meter will vary according to the model, the basics are the same. Sponsored Links Litmus paper Red, blue and neutral paper in books and reels manufactured in UK www.kaagat.com 3 Wash the electrode. The electrode is usually encased in a glass tube and can be washed by squirting it with distilled water. 4 Prepare the machine for calibration. This can be done by pressing the "cal" button on the machine and then immersing the electrode in one of the calibration solutions provided with the machine. These solutions are generally supplied at pH levels of 4, 7 and 10, and are in specially prepared bottles. The pH meter must be calibrated in two different solutions, but it doesn't matter which two are used. 5 Continue with the calibration. After pressing the "Cal" button, immerse the electrode in one calibration solution. Once the readout on the digital screen stabilizes, press the button marked "OK." Repeat this using a different solution (for example, calibrate first with the pH 7 solution, then calibrate again with the pH 10 solution). 6 Wash the......

Words: 332 - Pages: 2

Premium Essay

Performance Measurement

...Introduction When a procedure, a model, an approach, a case study, or other research is undertaken, generally some measure of the system's performance is presented and perhaps analyzed. A large number of different types of performance measures have been used to characterize systems, particularly production, distribution, and inventory systems. Such a large number of available performance measures makes performance measure selection difficult. Generally, performance measurement research focuses on analyzing performance measurement systems that are already in use, categorizing performance measures and then studying the measures within a category, and building rules of thumb or frameworks by which performance measurement systems can be developed for various types of systems. Beamon (1996) presents a number of characteristics that are found in effective performance measurement systems, and can therefore be used in evaluation of these measurement systems. These characteristics include: inclusiveness (measurement of all pertinent aspects), universality (allow for comparison under various operating conditions), measurability (data required are measurable), and consistency (measures consistent with organization goals). Besides analyzing the measures based on their effectiveness, benchmarking is another important method that is used in performance measure evaluation. Benchmarking can be useful in that it can serve as a means of identifying improvement opportunities. Camp (1989)......

Words: 2747 - Pages: 11

Free Essay

Determination of Equivalence Point Using Ph Titration of Potassium Hydrogen Phalate and 0.1 N Sodium Hydroxide with Phenolphthalein Indicator

... Department of Chemistry Determination of equivalence point using pH titration of Potassium Hydrogen Phalate and 0.1 N Sodium Hydroxide with phenolphthalein indicator By Lee Binks (12647892), Penny Jinks (12993456) and Hong Links 13504733) Date submitted: Sept 18th , 2013 Abstract- The experiment described elucidates the fundamental principle of potentiometric titrations and determination of equivalence point using pH titration of Potassium Hydrogen Phalate and 0.1 N Sodium Hydroxide with the use of phenolphthalein indicator. The present study revealed the equivalence point by using simple titration curve which was further confirmed by using first and second derivative plots and pKa value of the Potassium Hydrogen Phalate. Introduction- The pH meter measures the pH of a solution and provides a direct method of obtaining a titration curve which is a graph of measured pH values versus the volume of titrant added in milliliters. The equivalence point is the point at which a stoichiometrically equivalent amount of base has been added to the acid. It does not mean that pH will be necessarily 7. KHP is a monoprotic acid. The neutralization with NaOH takes place in a 1:1 ratio HOOCC6H4COOK (aq) + NaOH ( aq) ---> C6H4 ( COO)2 2- (aq) + K+ (aq) + Na+ (aq) The equivalence point occurs in the region where there is a relatively large change in pH with a relatively small change in volume on the titration curve. The steeper the curve the......

Words: 1572 - Pages: 7

Premium Essay

Measurement

...Vershire also has some issues on the performance measurement report and evaluation system. Besides the inefficient way to measure manufacturing efficiency, performances of plant managers are measured based on the metric that they cannot directly control, i.e. sales. Therefore the related compensation packages and promotions mislead and lack sufficient incentives to motivate plant managers. As illustrated before, it is inappropriate to treat a plant as a profit center, rather than an expense center. Due to the manufacturing nature, plants have fully control on costs and should be motivated to lower the cost to make contributions to the revenue. Monthly performance evaluation report and supporting report for a plant showing both sales and costs in Exhibit 2 and 3 are submitted by each plant and reviewed by corporate management. Items that failed to meet the budgeted targets require explanations from plant managers, no matter it is cost-related or sales-related. Moreover, sales report on plant basis is confusing by itself since products are sold on district basis. The sales department had sole responsibility for the price and delivery schedules, which are not identical in each order because of contingencies such as traffic, weather, and sick staff. Plant managers have no way to explain the detailed decision making process of sales department to provide a sufficient explanation to sales failures without asking sales department. Furthermore, second-hand information tends to be......

Words: 629 - Pages: 3

Premium Essay

Measurement of Poverty

...Measurement of poverty By Name: Course code+ Name: Professor: Institution: City and State: Date: Measurement of Poverty Over the past few decades, there have been dramatic socio-economic and environmental changes around the globe. This has affected phenomena such as growth of the population, rapid urban development, and increased levels of poverty (D’Ambrosio, Deutsch and Silber, 2011). It has also influenced the changes in climatic conditions, increased natural calamities which impact the socio-economic development in different parts of the world. As a result, more people have been exposed to the negative effects of the occurrence of different hazards (D’Ambrosio, Deutsch and Silber, 2011). This leaves the affected populations suffering in poverty, thus increasing their numbers. Increased levels of poverty led to the formation of policies by the United Nations to combat the rising cases of poverty with the aim of reducing the number of people living in poverty by halve before the year 2015. Formulation of the millennium development goals which features the quantitative objectives and indicators set to measure the progress towards the war on poverty eradication took course. To establish the level of poverty in the society, therefore, measurement should be performed severally using a number of available tools (Davidson and Duclos, 2000). This will aid in determination of the extent of its effects throughout the world. This paper aims at discovering the......

Words: 3454 - Pages: 14

Free Essay

Levels of Measurement

...Normally, when one hears the term measurement, they may think in terms of measuring the length of something (e.g., the length of a piece of wood) or measuring a quantity of something (ie. a cup of flour).This represents a limited use of the term measurement. In statistics, the term measurement is used more broadly and is more appropriately termed scales of measurement. Scales of measurement refer to ways in which variables/numbers are defined and categorized. Each scale of measurement has certain properties which in turn determines the appropriateness for use of certain statistical analyses. The four scales of measurement are nominal, ordinal, interval, and ratio. Nominal: Categorical data and numbers that are simply used as identifiers or names represent a nominal scale of measurement. Numbers on the back of a baseball jersey and your social security number are examples of nominal data. If I conduct a study and I'm including gender as a variable, I may code Female as 1 and Male as 2 or visa versa when I enter my data into the computer. Thus, I am using the numbers 1 and 2 to represent categories of data. Ordinal: An ordinal scale of measurement represents an ordered series of relationships or rank order. Individuals competing in a contest may be fortunate to achieve first, second, or third place. first, second, and third place represent ordinal data. If Roscoe takes first and Wilbur takes second, we do not know if the competition was close; we only know that Roscoe......

Words: 574 - Pages: 3

Free Essay

Milk Ph Level Comparison

...Ateneo de Naga University High School Km. 7, John Phelan Drive, Pacol, Naga City The Comparative Analysis of the pH Level and Lactose Content of Fresh Milks and Processed Powdered Milks Jenna Joyce Ebona Glowsy Cabral Nicole de Vera Rogelenne An Borja Renz John Alba Johannah Aguayo RESEARCHERS Mrs. Ma. Wendy Solomo ADVISER ACKNOWLEDGEMENT This study aimed to compare the pH level and Lactose content of fresh milks (Cow’s milk, Goat’s milk, and Carabao’s milk) and processed powdered milks (Bear Brand, Nido, and Alaska) in order to guide the consumers in choosing what kind of milk to buy. Nevertheless, this study would not be possible without the everlasting love, care, and assistance from the researcher’s family, who supported in the collection of milk samples for the experiment and encouraged the researchers when they were taken aback by discouragement. The researchers would never forget to acknowledge Mrs. Ma. Wendy Solomo who extended, in her precious time, patience in explaining the whole course of the research and gave words of encouragement and for the continuous help, guidance, and assistance extended to the researchers. Furthermore, the researchers would like to thank their friends especially their classmates for the shared information and for helping the researchers whenever they need help. The researchers would alsow like to thank the staffs at the INECAR......

Words: 3065 - Pages: 13

Premium Essay

Ph Reports

...Unit 15 P5 PH report Hypothesis: the aim of this experiment was to see the effects that different levels of PH values has upon amylase enzyme activity. Apparatus: * Stirring rod * Dimple tray * Five pipettes * Syringe * Starch solution * Amylase * Iodine * Buffer of PH4, 7 and 10 * Test tube holder * Stopwatch Procedure: the first thing to do would be to ensure that you have all of the equipment needs for the experiment. The first step is to get a test tube and place it within the test tube holder, once you’ve done that you get a pipette using the pipette take some of the iodine and then very carefully put to drops of the iodine solution into the test tube. After that gets another pipette using that one to very gently to store 2 mL of the amylase enzyme solution, then very carefully put to drops of the amylase solution into the test tube. Once you’ve done that you get a third pipette and use that one to store 1 mL of the PH4 buffer, you then very carefully put the 1 mL of the pH4 buffer into the test tube, the second to last thing that is now added is 2 mL of starch. Once all of these chemicals have been put into one test tube it is then gently stirred, your lab partner will at the same time put 2 drops of iodine into 4 wells of the dimple tray, labelled 4 wells as 10 seconds, 20 seconds, 30 seconds and 40 seconds, using a stopwatch keep an eye on the time and wants 10 seconds of past use a syringe get a few drops of your......

Words: 575 - Pages: 3

Free Essay

Potentiometric Measurement – Determination of the Ph of an Alkali Solution, a Buffer Solution , a Hair Shampoo, a Hand Wash and Distilled Water.

...Title Potentiometric Measurement – Determination of the pH of an alkali solution, a buffer solution , a hair shampoo, a hand wash and distilled water. Objective To carry out potentiometric measurement to determine whether a solution is controlled by the buffer system or by a strong electrolyte. To prepare the samples with different concentration by adding distilled water. Introduction The potentiometric measurement is to determine the difference in potential between a working (an indicator) electrode and a counter (reference) electrode by using the potentiometer. The working or indicator electrode represents the cathode which locates at the right half-cell whereas the counter or reference electrode represents the cathode which locate at the left half-cell. In most of the chemical processes, pH is one of the most general laboratory measurements because many chemical processes are dependent on pH. The speed rate of chemical reactions can often be significantly altered by changing the pH of the solution. Apart from that, the solubility of many chemicals in solution and their bio-availability is dependent on pH. The physiological chemistry of living organisms usually has very specific pH boundaries. The term pH derives from a combination of pH for the power and H for the symbol of the element Hydrogen. pH is the negative log of the activity of hydrogen ions. pH = 〖log〗_(10 a) H^+ At a given temperature, pH represents the activity of......

Words: 1088 - Pages: 5

Free Essay

Ph Experiment

...Neutrals pH Scale Investigations ___ By Dominic Krusniak Our Experiment We were challenged to test the accuracy of cabbage juice as an indicator for classifying liquid solutions as an acid, base, or a neutral. We believed that red cabbage juice would be an accurate indicator for classifying liquid solutions. We came to this conclusion because we had 2 reviewed other reports testing red cabbage juice in a similar way to our method with an outcome in agreement to our hypothesis. We tested our hypothesis by gathering two tablespoons of the liquids: dish soap, ammonia, baking soda, tea, pop, lemon juice, and vinegar. We poured them into separate, clean plastic cups. We scooped two tablespoons of the cabbage juice solution into each chemical and then we observed the color changes (see the data table below). Name of Chemical Dish Soap Ammonia Baking Soda Color Changed Green Green Blue Green Tea Pop Cream of Tatar solution Grey Pink Rose Pink Pink Lemon Juice Vinegar Dark Pink Dark Pink We claim that all of the chemicals which turned a green or blue color (i.e. dish soap, ammonia, and baking soda.) were bases. Therefore, we also said that all of the chemicals that turned pink or red, where acidic. (i.e. Tea,soda,cream of tatar solution,lemon juice, and vinegar.) The above mentioned facts support our hypothesis. In an article we read to begin our lab, we learned about pH and the pH scale. We read that pH was a ranking that classified things as acidic or basic. The......

Words: 469 - Pages: 2

Premium Essay

Ph Measurement and Buffer Preparation

...pH Measurement and Buffer Preparation Quilao, M.H., Regua, R., Reinoso, L.A., Rojas, K.J., Sabalberino, E.M. Abstract In this experiment, phosphate buffer solution was prepared and the pH of the buffer and samples were determined through different liquid indicators and the pH meter. Also, the buffer capacity of the prepared buffer solution was calculated. The group was tasked to prepare a 250 mL phosphate buffer solution of pH 8.0 using dihydrogen phosphate ion (H2PO4) and primary sodium phosphate monohydrate (NaH2PO4.H2O). With the aid of 6.0 M HCl and 6.0 M NaOH, the pH of the buffer was adjusted while being recorded by the pH meter until it reached the desired pH. Afterwards, the buffer solution was introduced to Colorimetric Determination which used acid-base indicators. The buffer solution changed to color yellow when Thymol and Methyl red were added, blue when Bromophenol blue and Bromocresol green were included, purple for Bromocresol purple, pink for Phenol red and Phenolphthalein, and orange for Methyl orange. Introduction The measurement of the low concentration of hydrogen ions present in any biological process is expressed as pH. pH is used to measure the acidity and alkalinity of a solution. pH involves ionic activity, making it difficult to accurately predict the pH value of a solution. The pH is also known to greatly affect our biological system and any large changes could be dangerous, which is why there is a buffer present within our......

Words: 971 - Pages: 4

Free Essay

Ph Meter

...pH meter: Vital tool for food processing A pH meter is defined as a device that measures the pH level of a liquid solution by means of knowing the voltage amount of two electrodes submerged in a solution. The use of this tool is never new to us. It is commonly utilized in science laboratory activities in school to know the acidity and basicity of household materials people see everyday. However, our perception of its application is only limited by the idea that pH meters are mere “laboratory equipment” used in school. Many overlook its importance especially in food production. People, who consume food everyday do not know that these “mere tools”, called pH meters, are vitally needed in the food processing industries. How are pH meters used in food processing? Knowing the pH level and monitoring it all throughout the process of food manufacturing is a critical step in maintaining the production of high-quality foods. Making sure that the proper pH range is achieved during the physical and chemical reactions undergone by the food in the whole duration of the process is vital. Improper interpretation of pH tests can cause a huge drawback in the production of food. Incorrect measurement of pH level in jelly production for an instance, affects the consistency of the jelly and thus the product quality as a whole. Yogurt production, as an example, is generally dependent on the pH level for if ever, the proper pH level is not reached, discoloration (Yogurt becoming yellow)......

Words: 575 - Pages: 3

Free Essay

Measurement

...gives the following: D z f − xB B xD − z f and / or = = F xD − xB F xD − xB (3) The equations (3) define the “cut”, i.e what percentage of the total feedflow exits the column as distillate and bottoms product for specified inlet and outlet concentrations. From equations (3) it is apparent that distillate (D) and bottoms flow (B) are related to top and bottom product compositions (xD and xB) and are therefore potential manipulated variables. As expected, changes in F and zf will also affect xD and xB. Around the condenser and accumulator assuming a total condenser, the material balances are: • the material balance: • the component balance: V n = D + Ln Vnyn = DxD + LnxD (4) (5) Lm = Vb + B Lmxm = Vbyb + BxB (6) (7) and around the reboiler, • the material balance: • the component balance: For a liquid feed Lm = Ln+ F Assuming that the molar flows of liquid and vapour are constant through the column (constant molal overflow) then, L = Ln = Ln+1……etc. V = Vn = Vn+1 = ….etc. Therefore: D = V - L and B = F + L - V (8) Equations (3) demonstrate that D and B may be used to regulated xD and xB, based upon the relationships (equation 8) it is obvious that L and V will also affect the product compositions. 3 Summary: the potential manipulated variables for product compositions are D, B, L and V. Column control strategies (an introduction) A ‘bottom – up’ approach should be adopted whereby variables that......

Words: 2663 - Pages: 11

Free Essay

3 Ph - 5 Ph

...motors are invariably supplied by ac/dc/ac converters. Thus, the focus of the research on the multiphase electric drive is limited to the modeling and control of the supply systems (i.e., the inverters [23]–[33]). Little effort is made to develop any static transformation system to change 3 and the phase number from three to -phase (where odd). The scenario has now changed with this paper, proposing a novel phase transformation system which converts an available three-phase supply to an output five-phase supply. Multiphase, especially a 6-phase and 12-phase system is found to produce less ripple with a higher frequency of ripple in an ac–dc rectifier system. Thus, 6- and 12-phase transformers are designed to feed a multipulse rectifier system and the technology has matured. Recently, a 24-phase and 36-phase transformer system have been proposed for supplying a multipulse rectifier system [34]–[37]. The reason of choice for a 6-, 12-, or 24-phase system is that these numbers are multiples of three and designing this type of system is simple and straightforward. However, increasing the number of phases certainly enhances the complexity of the system. None of these designs are available for an odd number of phases, such as 5, 7, 11, etc., as far as the authors know. 0885-8977/$26.00 © 2010 IEEE 1638 IEEE TRANSACTIONS ON POWER DELIVERY, VOL. 25, NO. 3, JULY 2010 Fig. 3. Phasor diagram of the proposed transformer connection (star-star). The usual......

Words: 4370 - Pages: 18

Free Essay

Measurement

...MEASUREMENT: Measurement is the heart of scientific discipline. It is the process of assigning numbers to objects or observations, the level of measurement being a function of the rules under which the numbers are assigned. According to Zikmund, “Measurement is the process of describing some property of a phenomenon of interest, usually by assigning numbers in a reliable and valid way”. In our daily life we are said to measure when we use some yardsticks to determine weight, height or some other features of a physical object. We also measure when we judge how well we like a song, intelligence, achievement, attitude, perception regarding quality. We, thus, measure physical object as well as abstract concepts. Measurement is a relatively complex & demanding task, especially so when it concerns qualitative or abstract phenomena. So, technically speaking, measurement is a process of mapping aspects of a domain onto other aspects of a range according to some rule of correspondence. It is a collection of quantitative data. TYPES OF MEASUREMENT: 1) Nominal measurement: All qualitative measurements are nominal, regardless of whether the categories differ from one another only in names. Nominal measurements represent the most elementary level of measurement in which values are assigned to an object for identification or classification purposes only. In nominal level of measurement, the categories differ from one another only in names. In other words, one category of a......

Words: 4319 - Pages: 18